Quick Detail
Place of Origin
HS-CODE
84-
Package & Delivery Lead Time
Detailed Description
Main Features
1.Four analog channels, 200MHz maximum bandwidth, 2GSa/s maximum real-time sample rate, 50GSa/s maximum equivalent sample rate
2.5.7 inch, QVGA (320× 240), 64K colors TFT LCD and LED backlight source technology enable the wave displays more vivid with lower power dissipation and longer life
3.Conform to LXI consortium instrument standard class C, which enable to create and reset testing system fast, economically and efficiently
4.Abundant trigger types: Edge, Pulse Width,Video, Pattern and Alternative triggers
5.Unique adjustable trigger sensitivity enables to meet different demands
6.Enable to measure 22 types of wave parameters and track measurements via cursor automatically
7.Unique waveform record and replay function
8.Fine delayed scan function
9.Built-in FFT function, hold practical digital filters
10.Pass/Fail detection function
11.Math operations available to multiple waves
12.Powerful PC application software UltraScope
13.Standard configure interface: USB Device,
14.Dual USB Host, LAN, support U disk storage and Pict Bridge print standard
15.Support for remote command control
Technical Specifications
Acquisition
Sample Modes
Real-Time Sample
Equivalent Sample
Sample Rate
2 GSa/s (half channel )
1 GSa/s (each channel)
50 GSa/s[2]
Averages
A waveform will be displayed one time while all the channels finish N times
Inputs
Input Coupling
DC, AC, GND
Input Impedance
1MΩ± 2.0%
The input capacity is 18pF± 3pF
Probe Attenuation Factors
0.001X, 0.01X, 0.1X, 1X, 2X, 5X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X
Maximum Input
Voltage
Maximum Input Voltage of the analog channel:
CAT I 300Vrms, 1000Vpk; transient overvoltage 1000Vpk
CAT II 100Vrms, 1000Vpk
RP2200 10:1, CAT II 300Vrms
RP3200 10:1, CAT II 300Vrms
RP3300 10:1, CAT II 300Vrms
Time Delay between
Channel (typical)
500ps
Horizontal
Sample Rate Range
3.65Sa/s-2GSa/s (Real-Time), 3.65Sa/s-50GSa/s (Equivalent-time)
Waveform
Interpolation
Sin(x)/x
Record Length
Up to 16k samples for half channel[1]
8k samples for each channel
Scanning Speed
Range
(Sec/div)
1ns/div~50s/div, DS1204B
2ns/div~50s/div, DS1104B
5ns/div~50s/div, DS1074B
1-2-5 Sequence
Sample Rate and
Delay Time Accuracy
± 50ppm (any time interval ≥1ms)
Delta Time
Measurement
Accuracy
(Full Bandwidth)
Single: ± (1 sample interval + 50ppm × reading + 0.6 ns)
>16 averages: ± (1sample interval +