Quick Detail
Place of Origin
HS-CODE
38-
Package & Delivery Lead Time
Detailed Description
Model:W277257
Chemistry analyzer can be used for routine clinical biochemistry, special protein, immune chemistry, drug testing, and so Determination.
Computer equipment DOC system using the latest technology to achieve the high reliability of the product. Rich applications software products so that the new user-oriented concept.
Equipment can only operate with 16 keys, and can be small keyboard and mouse operation, simple and clear, and easy to use. Show steps and the current state of Chinese language interface easy to learn.
Sample Testing: Power opened, as long as the movement of the cursor select projects, according to YES button, which entered beta status.
Parameter settings item analysis, quality control materials settings, the standard curve of the project are determined in accordance with the screen only tips, move the cursor keys can directly enter the working state.
File Management: Store up to 25,000 test results
Data recovery power
Can store 160 projects
-- Saved 45 commonly used parameters of the project.
-- Reagent to all users, (including storage project) users can set their own projects or modify parameters and permanent preservation.
Test the entire process
-- Can be immediately found unusual circumstances, such as bubble interference, in order to test-retest.
-- Dynamic test, automatically correct abnormal specimens, re-calculated results, and not have to re-determination.
Show Determination
End (EP), 2:00 (FT), dynamic Act (RA), dual-wavelength method of blank samples
Real-time quality control
-- Each of the projects can be set two quality control samples.
-- Can be given within 30 days of statistical quality control and quality control plans, image display or print the report.
-- Two-way RS232 communication interface with external computer communications.
Reliable calibration mode
-- Six standard curve
1:00 standards, 2:00 standard linear equations.
Multi-point standard (3-8) linear equations, nonlinear equations (polyline, conic, three curves, the logarithmic equation, three and four imitation-like function parameter equation).
Improve the printing function
Chinese character input patient information, print integrity of the integrated chemistry test report
-- Print report by project
-- Based on the sample, print report
-- Print report by the patient #
-- Print project parameters
Medical records: Print by Date (show) cases have been reported.
Case, by printing (show) cases have been reported.
Main technical parameters
Source: Halogen Tungsten Lamp (6 V10W)
Wavelength range :340-700 nm
Wavelength selection: - 9 wavelength automatic selection
-- Basic configurations: 7 wavelength (340405492510545578630 nm) can also be obtained in the following two wavelengths matching (380450650690 nm)
-- Spectral Bandwidth: ≤ 12 nm
Metering range 0-2.5 A
Metering resolution: - Show 0.001 Abs
-- Internal computer 0.0001 Abs
Precision: CV ≤ 1%
Stability: ≤ 0.00 2, A / h
Cross-contamination: ≤ 1.0%
Color plate: - Trace flows pool, 30 μ l capacity
-- Temperature: 25,30,37 ℃ temperature accuracy of 0.1 ℃
Sample volume: 500 μ l
Display: 10.4 "LCD Caixian (XD811L type)
15 "Caixian (XD811C type)
Keyboard: 16 Bond films touch button
Printers: an external 80 (A4) inkjet printer
Power: AC 220V/50Hz 150VA
Size: 450 * 420 * 280
Weight: 12 Kg (host)
Use of the environment: - room temperature :5-40 ℃
-- Relative Humidity: 10-93%