Model No : THS-900B ( 270L ) |
|
|
|
|
1 |
Internal Dimension:W900xH600xD500 mm |
|
|
External dimension:W1500xH1300xD1000mm |
|
|
|
|
|
|
|
|
|
2 |
Temperature Range: |
|
|
RT(+15℃)~ +75℃ Programmable |
|
|
|
|
|
|
|
|
|
3 |
Humidity Range: |
|
|
60%~95%R.H. Programmable or turn off this fuction |
|
|
|
|
|
|
|
|
|
4 |
Machine accuracy: |
|
1.Analytic accuraty: |
|
Temperature: 0.1 ℃ |
|
Humidity : 0.1 % |
|
2.Control accuracy : |
|
Temperature ± 0.5℃ |
|
Humidity : ±2.5%RH |
|
3.Distribution uniformity: |
|
Temperature:±2℃ |
|
Humidity:±3.5%(75%R.H. higher)±4.5%(75%R.H. lower) |
|
|
|
|
5 |
(A)Salt Spray Test: NSS,ACSS |
|
1.testing zone temperature : 35 Deg C (+/-) 2 Deg C |
|
2.Saturated air barrels temperature : 47 Deg C (+/-) 2 Deg C |
|
(B)Corrission resistance test: |
|
1.Laboratory: 50 Deg C (+/-) 2Deg C |
|
2.Saturated air barrels: 63 Deg C (+/-) 2 Deg C |
|
|
|
|
6 |
Air supply system: |
|
1.Air pressure is 1 kg/cm2, divided into two sections adjustment |
|
2.First step is rough adjustment : 2 kg/cm2 imported type air filter with drainage function, the second step is precision adjustment : 1 g/cm2 Pressure gauge |
|
|
|
|
7 |
Spray fog method: |
|
1.Adopt Knut principle to absorb brine to atomization, uniform atomization degree, to ensure continuous testing standard. |
|
2.Fog Spray quantity: adjustable type 1-2ml(ml/80cm2/h) (average quantity of 16 hours) |
|
|
|
3.Spray method: continuous spray. |
|
|
|
|
|
|
8 |
Heating system: |
|
Adopting the direct heating mode, fast heating, reduce the standby time, when the temperature reaches ,the ON/OFF switch shift automatically,temperature precision, less power consumption. |
|
|
|
|
9 |
Control System: |
|
1.Test Chamber temperature controller 0 ~ 99.9 Deg C , LCD display with Dual digital display, with P.T.D automatic calculation function, temperature deviation ± 0.3 Deg C |
|
2.Saturated air barrel temperature controller 0 ~ 99.9 Deg C , LCD display with Dual digital display, with P.I.D automatic calculation function, error of the temperature control in ±0.3 Deg C |
|
3.timmer : 0~9990hr |
|
|
|
4.Digital controller:0.1s~9990hr |
|
|
|
5. air pressure range of the gauge: 0~5kg/cm2 |
|
|
|
6.temperature senser |
|
|
|
7.electromagnetic valve |
|
|
|
8. electric relay |
|
|
|
|
|
|
11 |
Water system: |
|
adopting automatic adding water supplement system, automatically added, when the water level is too low |
|
|
12 |
Safety Protection Device: |
|
automatic power cut off supply device while water level is too low |
|
automatic power cut off supply device for over temperature |
|
alarm light device |
|
|
13 |
Accessories: |
|
sample Shelf/sample stick 4 pcs each |
|
1 pc brine mixing barrel; 1 pc standard measuring cup; HP test paper |
|
3.1 pc dry bulb cup |
|
4.1 pc 80cm2 collector |
|
1 pc glass spray nozzle |
|
1 pc Wet bulb cup |
|
1 set fog dispersal system |
|
1 pc meter regulator |
|
reagent Nacl 2 bottles (500 g/bottle) |
|
1 pc Air compressor ( Optional ) |
|
|
|
|
14 |
Power: AC380V , three phase 5 lines; 50/60Hz; 6 mm2 copper wire |
|
|
|
|
|
|
15 |
Delivery Attached Documents:Operation Manual,Warranty Card |
|